Nl-Fr

View abstract

This abstract is assigned to session AOB Free papers
TitleMulticenter defocus curve evaluation of a novel trifocal presbyopia correcting IOL – 6 months post-op results
Abstract Nr.2082
PurposeTo determine the Binocular Defocus Curve of a novel trifocal lens, the AcrySof IQ PanOptix IOL at 6 months post-op.
MethodsA prospective, single arm, non-randomized, multi-center study was conducted involving the bilateral implantation of AcrySof IQ PanOptix Presbyopia Correcting IOL in 151 subjects. The binocular defocus curves were evaluated under photopic lighting conditions.
ResultsA total of 148 subjects (38% M: 62% F) with a mean age of 68.9 ± 9.3 years were implanted bilaterally with the AcrySof IQ PanOptix IOL and had their binocular Defocus Curve assessed at 6-months post-op. The Mean Defocus curve VA from 0.00D to -3.00D ranged from 0.1 to 0.0 logMAR. Average VA at distance (0.00D), intermediate (-1.50D) and near (-2.50D) was 0.0 ± 0.1D, 0.0 ± 0.1D and 0.1 ± 0.1D, respectively.
ConclusionThe early results from this study show very good visual performance (VA better or equal to 20/25) of the AcrySof IQ PanOptix IOL across the whole range of defocus.
Conflict of interestYes
Details of conflicting interestsConsultant & study investigator Alcon-Novartis
Authors 1
Last nameSALLET
InitialsG
DepartmentOoginstituut
CityAalst
Authors 2
Last nameKOHNEN
InitialsT
DepartmentGoehte University
CityFrankfurt
top ^