View abstract

Cet abstract a été assigné à session OB Free papers: Anterior segment
TitreThe Zeiss VisuMax Femtosecond laser as a Refractive tool-A Paradigm Shift –FLEX-Excision instead of ablation.
Abstract Nr.260
ButTo discuss technical features and results of the Zeiss scientific study group and early personal experience using the- Zeiss VisuMax Femtosecond Laser - as a refractive tool.
MéthodesThe VisuMax Zeiss Femtosecond laser creates two very precise corneal cuts allowing FLEX: Femtosecond Lenticule Extraction for correction of refractive anomalies. With FLEX-instead of ablating tissue-tissue is excised with a precision of 4 micron in order to obtain the planned refractive change
RésultatsThe precision of the VisuMax equals the resolution of OCT measurement(4 Micron) .
At 12 months,in international and personal series, Refractive outcomes for 900 myopic eyes were 97%+/- 0, 5 d obtained- intended using the Flex technique. At 6 months ,98 % of eyes were 10/10 (20/20) uncorrected in both series.
Not one eye lost more than one line of Bscva. No complications were encountered . Patient satisfaction was very high. The stability over a 6 month fup was enormous with less than 0, 1 Dptr. Change.over 6 months.
ConclusionThe VisuMax Zeiss Femtosecond laser revolutionizes Refractive surgery with a new -extremely safe -and less invasive approach “FLEX -Excision instead of Ablation”.
Auteur 1
NomGOES
InitialesSr
InstitutGOES OOGCENTRUM
VilleANTWERPEN